rotor with straight cutting edges and fixed knives made of tool steel
+ sieve insert: 0.25 mm trapezoidal perforation stainless steel
Feed quantity:
60 g
Feed Size:
3 mm
Grinding time:
3 min
Final fineness:
< 0,25 mm
Comments:
Sound of grinding has been stopped after a grinding time of 180 seconds. Afterwards, an output of 45 g has been found in the collecting vessel.
Inside the grinding chamber, a higher amount of samples use to stick lightly by electrostatic charge. Because cleaning can be performed rapidly by vacuuming with a soft brush, and almost all openings of the sieve insert use to remain free, we are guessing that even higher amounts of sample can be ground this way.
For large amounts of sample, we recommend using the optional sample exhauster system with cyclone separator which is available for Universal Cutting Mill PULVERISETTE 19.