M0097010 | |
Chemistry | |
< 2mm | |
< 0,2mm | |
150g | |
Desired quantity can be ground with our Planetary Mono Mill PULVERISETTE 6 classic line as well as our Variable Speed Rotor Mill PULVERISETTE 14. |
Planetary Mono Mill PULVERISETTE 6 classic linemain disk speed: 600 rpm |
||
Equipment: |
500 ml grinding bowl made of agate (SiO2) + 10x 30 mm ø agate grinding balls |
|
|
150 g | |
|
2 mm | |
|
3 min | |
|
< 100 µm (fine powder) | |
|
Given sample use to fit inside of a 500ml grinding bowl.
After 3 minutes of dry grinding, sample has been ground to fine powder. No particles > 100µm have been found by optical microscopy afterwards. |
Planetary Mono Mill PULVERISETTE 6 classic linemain disk speed: 600 rpm |
||
Equipment: |
500 ml grinding bowl made of agate (SiO2) + 10x 30 mm ø agate balls |
|
|
150 g | |
|
2 mm | |
|
10 min | |
|
< 100 µm (fine powder) | |
|
Sample did not start sticking after 3 minutes of grinding. For this, we have ground the same kind of sample for totally 10 minutes. Still sample did not start sticking. For this, even longer dry grinding times are possible. Fineness did not change significantly compared to result 1 (after 3 minutes of grinding). Fineness has been determined my optical microscopy.
|
Variable Speed Rotor Mill PULVERISETTE 14 classic linespeed: 6.000 rpm |
||
Equipment: |
impact rotor with 12 ribs made of stainless steel + sieve ring: 0.5 mm trapezoidal perforation stainless steel |
|
|
80 g | |
|
2 mm | |
|
30 s | |
|
< 0,5 mm | |
|
A bit of fine dust uses to escape by the feeding funnel. By using the accessories for grinding large quantities, desired amount of 150g of sample can be ground in one batch and dust exportation will be minimized. |
Variable Speed Rotor Mill PULVERISETTE 14 classic linespeed: 6.000 rpm |
||
Equipment: |
impact rotor with 12 ribs made of stainless steel + 0.20 mm trapezoidal perforation stainless steel |
|
|
80 g | |
|
2 mm | |
|
30 s | |
|
<0,2 mm | |
|
A bit of fine dust uses to escape by the feeding funnel. By using the accessories for grinding large quantities, desired amount of 150g of sample can be ground in one batch and dust exportation will be minimized. |