Impact rotor with 12 ribs, made of stainless steel + 0,2 mm trapezoidal perforation sieve ring
Feed quantity:
15 g
Feed Size:
<5 mm*1
Additive:
liquid nitrogen
Grinding time:
4 min
Final fineness:
<200 µm
Comments:
*1:pre-cut with siccors.
15g is the maximum sample size that can be ground during one run, because of the warming of the accessories. A grinding without liquid nitrogen is not possible, because otherwise the sample would melt directly.
We recommend the use of a sieve ring with reinforced edges.